Display specifications
- Microprocessed Seven Segment LED display
- Microprocessed Monochromatic Graphic Display Monitor 5.5" (OPT)
- Microprocessed Colorful Graphic Display Monitor 8.4" (OPT)
Microprocessor control system
- Three modes of operation
- Skin or servo mode
- Manual mode
- Pre-heating mode
- Skin or servo mode control range - 34°C to 38°C
- Manual mode - Heater Output adjustable from 0-100% in 10% increments
- Pre heating mode - Fixed at 30%
- Auto test function performed at power on
- Temperature measurement accuracy - ±-0.2°C
- Skin probe accuracy - ±0.1°C (Range 25° to 45°C)
- Temperature display resolution - ±0.1°C
- Heating element - 560W
- Heater warmup time at 100% power less than 2 mins
- Auxiliary LED light
Physical characteristics
- Height - 1886mm to 2086mm (Optional)
- Depth - 1143mm
- Width - 1000mm
- Height -1800mm
- Mechanical Light adjustment of bed in 3 steps by technical position
- Two drawer storage system
- Bed tilting - ± 12°
- Continuous Electrical tilting - ± 12°(Optional)
- Mattress - 65 X 45 X 5cm
- Bearer Reel. for Wires
- Auxiliary tray
- Infusion pump stand
- IV pole
Weighing scale specifications (OPT)
- Range - 200 gms to 10kg
- Accuracy: -±4gm
T-piece resuscitator specifications (OPT)
- Manometer:-20 to 80cm 620
- Maximum Pressure - 70cm 620
- PIP - 5 to 70cm 820 @ 8LPM
- PEEP - 2 to 10cm H20 @ 8LPM
- Gas inlet floss range - SLPM (min) to 15LPM (max)
- 02 concentration - 21% to 100% depending on the inlet gas source
Air Venturi Suction (OPT)
- Range - 0 to 180mmHg
- Accuracy: ± 5%
Blender (OPT)
- Range - 21% to 100%
- Accuracy - ± 3%
Bilitron 3006 BTB (OPT)
- Integrated top surface phototherapy
- Five super LED's
- Irradiance - 30 to 40uw/crn2.nm
Bilitron Bed for Ampla
- Integrated under surface phototherapy 8 super LED
- Gel mattress
- Irradiance - 30 to 45 uw/cm'.nm
Pulse Oxymeter (OPT)
- Nelcore technology
- Measurement range - SPO2 : 21 to 100%
- Pulse Rate - 10 to 250BPM
- SPO2 Accuracy - From 70 to 100%,°2 digits
- Pulse rate
- With motion: 05 digits
Environmental conditions (OPT)
- Operating temperature -19°C to 30°C
- Operating humidity: 15% to 95% - No Condensative