Baby Warmer
Home › Products › Baby Warmer
Leading Manufacturers, Exporters, Wholesaler, Retailer and Trader of Ampala Baby Warmer, Duetto Incubator Warmer, Fanem Baby Warmer, Infant Baby Warmer, Infant Radiant Baby Warmer and Infant Radiant Warmer from Pune.
Display specifications
- Microprocessed Seven Segment LED display
- Microprocessed Monochromatic Graphic Display Monitor 5.5" (OPT)
- Microprocessed Colorful Graphic Display Monitor 8.4" (OPT)
Microprocessor control system
- Three modes of operation
- Skin or servo mode
- Manual mode
- Pre-heating mode
- Skin or servo mode control range - 34°C to 38°C
- Manual mode - Heater Output adjustable from 0-100% in 10% increments
- Pre heating mode - Fixed at 30%
- Auto test function performed at power on
- Temperature measurement accuracy - ±-0.2°C
- Skin probe accuracy - ±0.1°C (Range 25° to 45°C)
- Temperature display resolution - ±0.1°C
- Heating element - 560W
- Heater warmup time at 100% power less than 2 mins
- Auxiliary LED light
Physical characteristics
- Height - 1886mm to 2086mm (Optional)
- Depth - 1143mm
- Width - 1000mm
- Height -1800mm
- Mechanical Light adjustment of bed in 3 steps by technical position
- Two drawer storage system
- Bed tilting - ± 12°
- Continuous Electrical tilting - ± 12°(Optional)
- Mattress - 65 X 45 X 5cm
- Bearer Reel. for Wires
- Auxiliary tray
- Infusion pump stand
- IV pole
Weighing scale specifications (OPT)
- Range - 200 gms to 10kg
- Accuracy: -±4gm
T-piece resuscitator specifications (OPT)
- Manometer:-20 to 80cm 620
- Maximum Pressure - 70cm 620
- PIP - 5 to 70cm 820 @ 8LPM
- PEEP - 2 to 10cm H20 @ 8LPM
- Gas inlet floss range - SLPM (min) to 15LPM (max)
- 02 concentration - 21% to 100% depending on the inlet gas source
Air Venturi Suction (OPT)
- Range - 0 to 180mmHg
- Accuracy: ± 5%
Blender (OPT)
- Range - 21% to 100%
- Accuracy - ± 3%
Bilitron 3006 BTB (OPT)
- Integrated top surface phototherapy
- Five super LED's
- Irradiance - 30 to 40uw/crn2.nm
Bilitron Bed for Ampla
- Integrated under surface phototherapy 8 super LED
- Gel mattress
- Irradiance - 30 to 45 uw/cm'.nm
Pulse Oxymeter (OPT)
- Nelcore technology
- Measurement range - SPO2 : 21 to 100%
- Pulse Rate - 10 to 250BPM
- SPO2 Accuracy - From 70 to 100%,°2 digits
- Pulse rate
- With motion: 05 digits
Environmental conditions (OPT)
- Operating temperature -19°C to 30°C
- Operating humidity: 15% to 95% - No Condensative
Read More...
Features
- Microprocessed Seven Segment LED display
- Microprocessed Monochromatic Graphic Display Monitor
- Microprocessed Colorful Graphic Display Monitor 8.4"
Control system
- Three modes of operation
- Skin or servo mode
- Pre-heating mode
- Skin or servo mode control range: 34°C to 38°C
- Manual mode - Heater Output adjustable from 0-100% in
- Pre heating mode - Fixed at 30%
- Auto test function performed at Power on
- Temperature measurement accuracy - ±0.2°C
- Skin probe accuracy - ±1°C (Range 25" to 45°C)
- Temperature display resolution - ±0.1°C
- Heating element - 560W
- Heater warmup time at 100% power in less than 2 mins
- Auxiliary LED light
Physical characteristics
- Height - 1886mm to 2086mm (Optional)
- Depth - 1143rnrnWidth: 1000mm
- Two drawer storage system(Optional)
- Continuous Electrical tilting: t 12"(Optional)
- Mattress: 65 X 45 X 5cm
- Infusion pump stand
Weighing scale specifications
- Range - 200 gms to 10kg
- Accuracy: ±4gm
Bilitron Bed for Ample
- Integrated under surface phototherapy
- Gel mattress
- 8 super LED's
- Irradiance: 30 to 45 uw/cninm
T-piece resuscitator specifications (Optional)
- Manometer -20 to 80cm H2O
- Maximum Pressure - 70cm H2O
- PIP - 5 to 70cm H2O @ 8LPM
- PEEP - 2 to 10cm H2O @ 8LPM
- Gas inlet flow range - 5LPM (min) to 15LPM (max)
- 02 concentration - 21% to 100% depending on the inlet gas source
Blender (Biomed.USFD A approved) (Optional)
- Range: 21% to 100%
- Accuracy, ± 3%
- Minimum input gas supply pressure - 25PSI
Bilitron 3006 BTB(Optional)
- Integrated top surface phototherapy
- Five super LED's
- Irradiance 30 to 40uw/cni nm
Read More...
Features:
- Micro Controller based Electronics
- Controls - Servo Skin/ Apgar
- Pre-warm - Use the equipment with 30% Heater Output in Manual Mode.
- Programmable Mute time
- No need of calibration required for temperature probe.
- High Intensity Quartz Heating element for effective Radiation (Ceramic heater Optional)
- Halogen Observation Lamp
- The Heater Module rotates 90° to both sides to facilitate
- X-Ray Procedures.
- The Parabolic Reflector is designed to send uniform heat to the bed surface.
- Memory Backup to restore set temperature and previous mode, in case of power failure.
- Optimum Bed dimensions ensure optimum work space for the nursing staff.
- Comprehensive Alarms and safety cut-offs.
- Alert (Alarm) when sensor is detached from the patient's skin (optional)
- Instant Warming and Uniform heating
Controller
- Electronics : High speed micro controller
- Mode of operation : Servo skin, manual & apgar
- Skin mode control range : 30 Deg. C to 38 Deg. C
- Manual mode timer : 1 min to 15 mins (selectable), 15 mins to 30 mins (optional)
- Apgar timer : Alerts at 1, 5, 10, 15 & 20 mins (selectable)
- Percentage of heater output : 10% to 100% in 10% increments
- Memory back up : Equipment restores the previous settings
- Programmable mute time : User can select the mute time
- Safety heater cut off : If the skin sensor is removed from the baby's skin
- Over temperature heater cut off : If the skin temperature is >39 Deg. C in any skin/manual mode
- Temperature sensor (probe) : Thermistor sensor
- Accuracy of skin sensor : ±0.1 C
- Accuracy of skin probe Interchangeability : 0.2 C
- Calibration of skin sensor (probe) : No need of calibration after Changing new probe
- Heater element : Quartz Infrared heater Ceramic heater (optional)
Other Details
- Power rating : 230V AC, 50Hz, 700 W
- Maximum heater wattage : 600 Watts
- Maximum power requirement : 700 Watts
- Examination light : 12V, 50W Halogen lamp
- X-ray facility : Underneath the bed
- Mattress tilt : ±30 degrees on both sides
- IV pole : Maximum load 2.2 Kg
- Pump mounting pole : Maximum load 2.2 Kg
- Castor : 4” castors 4 Nos. (2 with brake & 2 without brake
- Monitor shelf : 2 Nos. Fixed
- Height adjustable heater module : Available only in nice 2010
- Cabinet : Available only in nice 2010BC
- Operating noise level : <50 dBA
Controller Alarms:
- Check patient If the skin sensor is removed from
- the baby's skin
- Skin temperature high > 1.0 C from the set temperature
- Skin temperature low < 1.0 C from the set temperature
- Skin temperature probe fail Probe short or open
- Over temperature If skin temperature is 39 C
- Heater fail Heater defect
- System fail Electronic component defect
- Power fail AC power interruption
Environmental
- Operating temperature - 15° C to 35° C
- Storage temperature - -20° C to 60° C
- Operating humidity - 5% to 90% RH, non condensing
- Storage humidity - 0% to 99% RH, non condensing
- Altitude - Sea level to 1-9 miles, 3 Km
- Quality - ISO 9001:2000 & ISO 13485:2003
- Electrical & product safety - IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-21
Read More...